作 者:韦冠一[1] 翟利华[1] 张子斌[1] 王长海[1] 李雪松[1] 万可友[1]
机构地区:[1]西北核技术研究所,陕西,西安,710024 西北核技术研究所,陕西,西安,710024 西北核技术研究所,陕西,西安,710024 西北核技术研究所,陕西,西安,710024 西北核技术研究所,陕西,西安,710024 西北核技术研究所,陕西,西安,710024
出 处:《质谱学报》2005年第z1期99-100,共2页Journal of Chinese Mass Spectrometry Society
摘 要:The simulation results of MAT-261 type ion source and ion optics with the SIMION 3D 7.0 software were given. By means of simulation, the ion's trajectory was studied, the cause of low abundance sensitivity was analyzed and the optimal poles' voltage ranges of the ion optics were found. Modifications were made to the LRIMS3000 type mass spectrometer according to result of the simulation. The abundance sensitivity of the mass spectrometer is improved to 2×10-6 from original 5×10-5.
关 键 词:ion-optics simulation ABUNDANCE sensitivity MASS SPECTROMETER