基于开孔球面高强度聚焦超声阵列的相控精度设计  被引量:2

Phased Accuracy Design Based on a Holed Spherical High Intensity Focused Ultrasound Array

在线阅读下载全文

作  者:陈琦[1] 张■ 陈亚珠[1] 

机构地区:[1]上海交通大学生物医学仪器研究所,上海200030

出  处:《上海交通大学学报》2003年第z1期95-98,共4页Journal of Shanghai Jiaotong University

摘  要:研究了超声换能器阵列驱动信号的相位调控精度对阵列聚焦声场的影响.基于开孔球面高强度聚焦超声(HIFU)阵列,提出了一种阵列驱动信号的5位量化精度相控方案,解决了相控HIFU治疗系统中驱动信号相位的非连续性调控问题.仿真结果表明,此种相控方案既使阵列声场保持了高声强的聚焦点,又维持了原有的栅瓣抑制能力,从而为相控HIFU阵列驱动系统的设计提供了可靠依据.The effect of phased accuracy of the driving signals for array transducers on focused ultrasound field was studied. Based on a holed spherical high intensity focused ultrasound(HIFU) array, a phased project of 5bit quantized accuracy for driving signals was suggested. It solves the problem of discrete adjust and control for driving signals in the phased HIFU treatment system. The simulation results show that this phased project can keep higher sound intensity and initial abilities of restraining the grating lobes of ultrasound field. It provides reliable foundation for the design of the phased HIFU driving system.

关 键 词:高强度聚焦超声 球面阵列 相控精度 

分 类 号:TP311[自动化与计算机技术—计算机软件与理论]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象