Brachial access technique for aortoiliac stenting revisited  被引量:1

Brachial access technique for aortoiliac stenting revisited

在线阅读下载全文

作  者:Gianluca Rigatelli Paolo Cardaioli Fabio dell'Avvocata Massimo Giordan 

机构地区:[1]Cardiovascular Invasive Diagnosis and Interventions Unit, Rovigo General Hospital, Rovigo, Italy

出  处:《Journal of Geriatric Cardiology》2007年第2期78-79,共2页老年心脏病学杂志(英文版)

摘  要:We report a modified technique to perform iliac artery stenting through the brachial artery access. A 6F Brite tip sheath (Cordis, Jonhson & Jonhson Medical, Miami Lakes, FL, USA) is inserted into either brachial artery and a standard 4F Judkins Right diagnostic catheter was inserted over a 260 cm 0.038” Terumo Stiff wire (Terumo Corp, Tokyo, Japan) through the sheath. The catheter is navigated down to the aortic bifurcation, and after selecting the common iliac artery ostium, the wire is navigated through the lesion and advanced to the ipsilateral superficial femoral arteries. The catheter should be then moved forward over the wires beyond the lesion and the Terumo guidewire is replaced by two 0.038” 260 cm Supracor wires (Boston Scientific Corporation, San Jose, CA, USA). In order to facilitate advancement of the stent without risk of dislodgement as well as to check the position with low contrast dose injection, a 6 F (or 7F if large stent is selected) 90cm Shuttle Flexor introducer long sheath (Cook Group, Bloomington, IN, USA) should be advanced over the Supracor wire until it reaches the common iliac artery ostium. A road-map technique can be used to check the ostium position in order to properly deploy the selected stent. This technique promises to be safe and effective offering more support than guiding catheter technique; moreover it reduces the stress on the arterial vessel at the subclavian site and enables a stiff balloon or stent catheter to be advanced even through a very elongated and calcified aorta without the risk of stent dislodgement.We report a modified technique to perform iliac artery stenting through the brachial artery access. A 6F Brite tip sheath (Cordis, Jonhson & Jonhson Medical, Miami Lakes, FL, USA) is inserted into either brachial artery and a standard 4F Judkins Right diagnostic catheter was inserted over a 260 cm 0.038” Terumo Stiff wire (Terumo Corp, Tokyo, Japan) through the sheath. The catheter is navigated down to the aortic bifurcation, and after selecting the common iliac artery ostium, the wire is navigated through the lesion and advanced to the ipsilateral superficial femoral arteries. The catheter should be then moved forward over the wires beyond the lesion and the Terumo guidewire is replaced by two 0.038” 260 cm Supracor wires (Boston Scientific Corporation, San Jose, CA, USA). In order to facilitate advancement of the stent without risk of dislodgement as well as to check the position with low contrast dose injection, a 6 F (or 7F if large stent is selected) 90cm Shuttle Flexor introducer long sheath (Cook Group, Bloomington, IN, USA) should be advanced over the Supracor wire until it reaches the common iliac artery ostium. A road-map technique can be used to check the ostium position in order to properly deploy the selected stent. This technique promises to be safe and effective offering more support than guiding catheter technique; moreover it reduces the stress on the arterial vessel at the subclavian site and enables a stiff balloon or stent catheter to be advanced even through a very elongated and calcified aorta without the risk of stent dislodgement.

关 键 词:BRACHIAL access PERCUTANEOUS ANGIOPLASTY STENT ENDOVASCULAR peripheral vascular disease 

分 类 号:R543[医药卫生—心血管疾病]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象