Z-Pinch软X射线探测器件的同步辐射标定  

Synchrotron Radiation Calibration for Soft X-ray Detector(In Chinese)

在线阅读下载全文

作  者:甯家敏[1] 郭存[1] 徐荣昆[1] 蒋世伦[1] 许泽平[1] 陈进川[1] 夏广新[1] 薛飞彪[1] 秦义[1] 

机构地区:[1]中国工程物理研究院核物理与化学研究所,绵阳621900

出  处:《中国核科技报告》2009年第1期71-77,共7页China Nuclear Science and Technology Report

摘  要:介绍了利用同步辐射装置3W1B光站软X光源标定X光胶片、闪烁体、透射光栅的实验。实验中标定了闪烁体的能量响应、透射光栅的透过率和衍射效率,标定出软X光胶片黑密度曲线和能量响应曲线。实验结果可用于快过程Z-Pinch X光辐射特性的诊断,如软X射线功率仪、光栅谱仪、针孔相机、一维时空成像系统等测试项目,保证了Z-Pinch实验数据的精度。The calibration experiments were carried out to X-ray film,scintillator and transmission grating by employing the soft X-ray station at 3W1B beam-line in Beijing synchrotron Radiation Facility.The experiments presented the black intensity curve and energy response curve of soft X-ray film.And the experimental results can be used in diagnosis of X-ray radiation characterization of Z-pinch,such as in the measurement of soft X-ray Power Meter,grating spectrometer,pinhole camera and one-dimension imaging system...

关 键 词:同步辐射 Z箍缩 软X射线 

分 类 号:O434.12[机械工程—光学工程]

 