用近红外光谱检测人体组织氧含量  被引量:16

Measuring human tissue oxygenation with near infrared spectroscopy

在线阅读下载全文

作  者:腾轶超[1] 丁海曙[1] 田丰华[1] 黄岚[1] 

机构地区:[1]清华大学医学院生物医学工程系,北京100084

出  处:《清华大学学报(自然科学版)》2004年第6期847-851,共5页Journal of Tsinghua University(Science and Technology)

基  金:国家自然科学基金资助项目 ( 697780 2 4) ;国家教育振兴计划项目 ( Jc2 0 0 10 2 2 )

摘  要:为防止病人因组织缺氧而产生各种临床并发症 ,需要检测病人的组织氧。红外组织氧含量检测技术是一种新的无损检测技术 ,相对于脉搏氧饱和度检测 ,可以检测局部人体组织的氧合状况。用自行设计的近红外组织氧含量检测仪 ,进行了前臂阻断实验、婴儿脑氧实验和肌氧实验。检测中考虑了近红外传感器与人体的耦合 ,合理选择了光源与检测器的距离以及检测器的数目。用该仪器测量结果与生理变化相符。结果表明Tissue oxygenation of patients need be measured and monitored to avoid complications caused by hypoxia in medical procedures. A new noninvasive technique was introduced to measure tissue oxygenation using near infrared spectroscopy (NIRS). The method gives better results than measurements of pulse oxygenation. A tissue oximeter was developed using NIRS. Experiments of forearm occlusion, pediatric brain oxygenation and muscle oxygenation were made by the oximeter. The coupling between the near infrared sensor and human body was in corporated into the system, and the distances between the light source and the detectors as well as the number of detectors were optimized. The results were consistent with physiological changes. Human tissue oxygenation can be noninvasively measured real time by NIRS.

关 键 词:近红外 组织氧 无损检测 

分 类 号:TN21[电子电信—物理电子学]

 