Cobalt Determination by Means of Flow Injection On-line Sorbent Preconcentration-Flame Atomic Absorption Spectrometry with 1-Nitroso-2-naphthol as a Complexing Reagent  

Cobalt Determination by Means of Flow Injection On-line Sorbent Preconcentration-Flame Atomic Absorption Spectrometry with 1-Nitroso-2-naphthol as a Complexing Reagent

在线阅读下载全文

作  者:YIN Xue-feng ALI Akbar XU Guang-ming LING Yun-yang GAO Jian YE Mei-ying 

机构地区:[1]Department of Chemistry,Zhejiang University,Hangzhou 310027,P.R.China

出  处:《Chemical Research in Chinese Universities》2001年第3期264-270,共7页高等学校化学研究(英文版)

基  金:Supported by Zhejiang Natural Science Foundation of China.

摘  要:The suitability of 1-nitroso-2-naphthol(NN) as a complexing agent for on-line preconcentration of cobalt eluted on the C_(18) microcolumn by means of the FI-FAAS system was tested. Various parameters affecting the complex formation and its elution were optimized. A 2.3×10^(-3) mol/L reagent solution and the aqueous sample solution acidified with 0.1% (volume fraction) nitric acid were on-line mixed (6.4 mL/min) on a reaction coil set at (65±1)℃ and flowed through the microcolumn for 30 s. The pH of the mixed solution was adjusted to 3—4 with HNO_3(1 mol/L) or NaOH(1 mol/L). The adsorbed complexes in the microcolumn were eluted into the nebulizer of FAAS in 10 s with ethanol acidified with 1% HNO_3(3.0 mL/min). A good precision(1.6% for 100μg/L Co(Ⅱ), n=10), a high enrichment factor 17.2, with detection limit (3σ) 3.2μg/L, and sample throughput (90 h^(-1)) were obtained. The method was applied to the certified reference materials(CRMs), NBS-362 and NBS-364, for the determination of cobalt and the results were in good agreement with the certified values.The suitability of 1-nitroso-2-naphthol(NN) as a complexing agent for on-line preconcentration of cobalt eluted on the C_(18) microcolumn by means of the FI-FAAS system was tested. Various parameters affecting the complex formation and its elution were optimized. A 2.3×10^(-3) mol/L reagent solution and the aqueous sample solution acidified with 0.1% (volume fraction) nitric acid were on-line mixed (6.4 mL/min) on a reaction coil set at (65±1)℃ and flowed through the microcolumn for 30 s. The pH of the mixed solution was adjusted to 3—4 with HNO_3(1 mol/L) or NaOH(1 mol/L). The adsorbed complexes in the microcolumn were eluted into the nebulizer of FAAS in 10 s with ethanol acidified with 1% HNO_3(3.0 mL/min). A good precision(1.6% for 100μg/L Co(Ⅱ), n=10), a high enrichment factor 17.2, with detection limit (3σ) 3.2μg/L, and sample throughput (90 h^(-1)) were obtained. The method was applied to the certified reference materials(CRMs), NBS-362 and NBS-364, for the determination of cobalt and the results were in good agreement with the certified values.

关 键 词:Cobalt FI on-line preconcentration AAS 1-Nitroso-2-naphthol 

分 类 号:O614.812[理学—无机化学]

 