椎弓根钢板治疗胸腰椎骨折存在的问题(附90例分析)  被引量:41

The Problems in the Fracture of Thoraco-Lumbar Spine Treated by the Plate Screwed in Vertebral Pedicle

在线阅读下载全文

作  者:董天华[1] 唐天驷[1] 郑祖根 沈忆新[2] 成茂华[2] 华金明 

机构地区:[1]苏州医学院附属第一医院骨科 [2]苏州医学院附属第二医院骨科,215004

出  处:《骨与关节损伤杂志》1993年第1期16-18,共3页The Journal of Bone and Joint Injury

摘  要:椎弓根钢板系改良Roy—Camille钢板,作者报告90例治疗经验,具有手术创伤小,复位满意,内固定坚强,安全等优点。但存在轴向撑开力不足,如对T_(8-10)、L_(4-5)骨折复位不满意,易发生螺钉折断、松动、椎弓根钻孔位置不当及继发性畸形等问题。本文分析了存在问题,并提出防治措施。The plate screwed in vertebral pedicle (P.S.P)was a modified Roy-Camille plate. The author present operative results in 90 cases and satisfactory reduction. Strong internal fixation. safety and few complication were its proper advantage.but the force of axial distraction is not enough. The problems were demonstrated in screw broken,loosen ,secondary deformity and the complete reduction was difficult to be obtained in T8-10and L4_5 segment.

关 键 词:椎弓根钢板 胸腰椎骨折 并发症 

分 类 号:R683.205[医药卫生—骨科学]

 