摘 要:目的 分析体部X -刀治疗过程的质量保证和质量控制。方法 采用标准电离剂量法和胶片剂量法对体部X -刀照射的剂量、摆位精度、靶点位置精度进行测量。结果 体部X -刀治疗最大误差为 3 % ,完全符合国家规定 5 %的标准。结论 QA和QC是保证体部XObjective To analyse quality assurance(QA) and quality control (QC) during the process of X-ray stereotactic treatment.Methods The dosage of body X-ray irradiation,the treatment room of linear accelevator and the position of targe were measured using the standard inonizing and film dosage methods.Results The maximum error rate of body X-ray irradiation treatment system was 3%,it was in keeping with the national standard level which is 5%.Conclusion It is an essential to establish a comprehensive QA and QC program to guarantee a good treatment precision of X-ray stereotactic irradiation.