医疗光学CT成像前期研究——相位调制仪测量血氧饱和度的模拟生物实验  被引量:1

The Prophase Research on Medical Treatment Optics CT Imaging——Experimentation of Phase Modulation Device Measure Hererogeneous Saturation on Simulate Biology

在线阅读下载全文

作  者:郭海峰[1] 杨康[1] 

机构地区:[1]沈阳工业学院信息科学与工程学院,辽宁沈阳110168

出  处:《中国医学物理学杂志》2004年第5期300-301,共2页Chinese Journal of Medical Physics

摘  要:目的:为光学CT的成像提供物理基础。材料:用相位调制仪测量直径为10cm,高度为12cm的塑料杯中的混合溶液。在该溶液中,选用0.1%~1%的Intralipid溶液作为散射物质,选用新鲜鸡血作为吸收物质,通过往溶液中加入氧气或酵母以改变血氧饱和度,观察相角的变化。结果:相位调制系统的实验结果与理论上是相符的。结论:相角算法能够保证计算精度。Objective: The physics base of optics CT imaging is afforded. Material: the blend liquor in plastic cup which diameter is 10 cm and highness is 12 cm i s measured by the phase modulation device. In this liquor, 0.1%~1% Intralipid l iquor serve as scatter substance and verdure chook blood serve as absorb substan ce, the change of the phase angle is obsevred while infusing oxygen or yeast int o this liquor for changing hererogeneous saturation. Result: It shows that, the experimentation result of PMS is in accord with the theory. Conclusion: The prec ision of computation algorithmic can be ensured by phase angle algorithmic.

关 键 词:相位调制装置 血氧饱和度 光学CT 成像 

分 类 号:TN911.73[电子电信—通信与信息系统]

 