出  处:《中国消毒学杂志》2004年第4期318-319,共2页Chinese Journal of Disinfection

摘  要:为测定器械消毒液中苯扎溴铵的含量 ,用二阶导数光谱法进行试验观察。以 2 6 6 .3和 2 6 8.8nm处峰、谷之间的振幅值作为苯扎溴铵的定量信息。结果 ,在溶液中苯扎溴铵≤ 30 0 μg/ml时 ,其含量与对应的吸收度呈线性关系 ,测定结果的相对标准偏差≤ 1.0 0 %。结论 ,该方法简便、快速 ,适用于医院制剂的定量分析。In order to measure bromogeramine content of the instrument disinfectant solution, second derivative spectroscopy was used for experimental observation. The amplitude values between peak and trough at 266.3 and 268.8 nm were used as the quantitative information of bromogeramine. The results showed that when bromogeramine content of the solution was ≤300μg/ml, the content had linear correlation with the corresponding absorbance and the relative standard deviation of the results measured was ≤1.00%. Conclusion: This method is simple and quick and is suitable for quantitative analysis of hospital preparations.

关 键 词:苯扎溴铵 器械消毒液 二阶导数光谱法 含量 医院制剂 试验观察 吸收度 线性关系 测定结果 溶液 