符合线路显像与PET显像中SUV的比较研究  被引量:11

Comparison of dual-head coincidence imaging SUV with PET imaging SUV: a phantom study

在线阅读下载全文

作  者:耿建华[1] 陈英茂[2] 陈盛祖[1] 田嘉禾[2] 何义杰[2] 乔士珍[1] 

机构地区:[1]中国医学科学院,中国协和医科大学肿瘤医院核医学科,北京100021 [2]解放军总医院核医学科

出  处:《中华核医学杂志》2004年第5期308-309,共2页Chinese Journal of Nuclear Medicine

摘  要:目的比较符合线路显像标准摄取值(SUV)与PET显像的SUV。方法用双探头符合显像仪及PET对模型显像,分别采用不同的重建算法重建,测定图像上热灶的SUV。结果对直径小于30mm热灶,相同大小时,PET得到的SUV高于符合线路显像;无论对PET还是符合线路显像,随热灶大小增加SUV增加;SUV与重建算法有关;选取的感兴趣区(ROI)越大,获得的SUV越小;由PET图像获得的热灶SUV可见,当热灶大于2倍的系统分辨率时,SUVmax接近热灶的真实值(SUVtrue)。结论符合线路显像的SUV低于PET显像;病灶大小、重建算法、ROI大小均影响SUV。Objective To compare dual-head coincidence imaging standard uptake value (SUV) with PET imaging SUV. Methods The scans were performed with dual-head coincidence imaging equipment and PET on a phantom which consisted of some hot lesions. The images were reconstructed with different algorithms. The hot lesions' SUVs were measured. Results For the same hot lesions, the SUV from PET images was larger than that from the dual-head coincidence images. For both the PET image and dual-head coincidence image, the SUV increased with the size of the lesions. The SUV was related with reconstruction algorithm. The larger the region of interest (ROI), the smaller the SUV of a lesion. The SUV from PET images showed the maximum SUV and was nearly coming to the true SUV value of the hot lesions 2 times larger than that of system resolution. Conclusion The SUV in dual-head coincidence imaging is lower than that in PET imaging; and the lesion size, reconstruction algorithm, and the size of ROI can influence the SUV.

关 键 词:符合线路显像 PET显像 SUV 比较研究 

分 类 号:R817[医药卫生—影像医学与核医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象