血液葡萄糖测定——葡萄糖氧化酶与已糖激酶之实验比较  

Determination of Glycemia-A comparision between Glucose-Oxidase and Hexokinase

在线阅读下载全文

作  者:杨明容[1] 张秋霞[1] 

机构地区:[1]昆明医学院检验科

出  处:《昆明医学院学报》1993年第1期44-46,共3页Journal of Kunming Medical College

摘  要:国产葡萄糖氧化酶(GOD)与进口已糖激酶(HK),GOD药盒进行实验对比,其特异性、准确性、精密度都接近HK的参考方法。国产GOD批内CV1.06%,批间CV1.55%,平均回收率96.7%-105.2%。血糖浓度在2.8-28mmol/L线性关系良好。与TRACE-HK比较r=0.999,与CORN-ING-HK比较r=0.997,与IL-HK比较r=0.996。141例健康正常人血糖浓度为4.84±0.58mmol/L。Domestic Glucose-Oxidase (GOD) has been compared with imported Hexokinase (HK) in the test of glucose of human plasma. The result indicates that both are similar in accuracy and precision according to the referential method of HK. The intra-and inter-batches CV of GOD are 1.55% and 1.06% respectively. Its recovery rate is 96.7~105.2% while the concen- tration of glycemia is 2.8~28.0mmol/L, which indicates a good linear correlation. In the comparision of Trace-HK, Corning-HK, 1L-HK and corning-GOD, the correlation coefficients, r, the order is as follows: 0.999, 0.997, 0.996, and 0.996. 141 healthy people have been examined fer glycemia. Their normal values are 3.68~6.00mmol / L.

关 键 词:葡萄糖氧化酶 己糖激酶 血液 

分 类 号:R446.112[医药卫生—诊断学]

 