交叉污染对血清甘油三酯自动分析的影响  被引量:7

Effects of cross-contamination on automated analysis of serum triglycerides

在线阅读下载全文

作  者:尤昕[1] 任淑子[1] 魏学峰[2] 金幸[1] 李玉花[1] 

机构地区:[1]延边大学医学院附属医院检验科,吉林延吉133000 [2]延吉市医院,吉林延吉133000

出  处:《吉林医学》2004年第10期20-21,共2页Jilin Medical Journal

摘  要:目的:了解自动生化仪测定中,交叉污染对血清甘油三酯测定的影响。方法:使用日立7600-010自动生化分析仪,先测1次已设置的所有项目,在每一个项目测试后连续测3次无机磷,以确定对无机磷交叉污染的项目。结果:胆汁酸、肌酸激酶MB型同工酶等试剂对无机磷测定结果有明显的影响。结论:自动生化分析仪测定血清甘油三酯存在项目间的交叉污染,有必要调整项目的通道以避免交叉污染。Objective To determine the effects of cross-contamination on the automated analysis results of serum Triglycerides.Method After each test that set up at Model7600-010automatic analyzer was detected once,Triglycerides test was continuously de-tected3times,to find out the tests that creating cross-contamination on the results of Triglycerides.Results The reagents of TBA and CK-MB significantly affected the results of Triglycerides.Conclusion It is necessary to change tests channel to prevent the effects of cross-contamination on automatic detecting of Triglycerides.

关 键 词:交叉污染 血清甘油三酯 自动生化分析仪 影响 自动分析 自动生化仪 胆汁酸 无机磷 同工酶 使用 

分 类 号:R446.1[医药卫生—诊断学]

 