作 者:冯彦林[1] 贺小红[1] 黄克敏[1] 余丰文[1] 刘德军[1] 袁建伟[1] 袁白虹[1] 苏少弟[1]
机构地区:[1]广东省佛山市第一人民医院PET/CT中心,广东佛山528000
出 处:《中国医学影像技术》2005年第1期127-130,共4页Chinese Journal of Medical Imaging Technology
摘 要:目的 探讨显像时的活性水平与显像时间对PET/CT图像质量的影响。方法 分别在Jaszczak模型内高活性短采集时间、推荐活性和采集时间以及低活性长采集时间下 ,以全身显像方式采集发射数据 ;采用X线CT作衰减校正(CTAC)和全身显像方式进行图像重建。相同患者按 5 .18MBq/kg给药 ,分别于给药 5 0min和 90min后采集数据 ,采用CTAC和全身显像方式进行图像重建。以 15 2 .0 2 6MBq、每床位采集 3min 3 0s对 1例 74kg女性志愿者进行显像。 结果 三种显像条件所得图像都可清晰分辨模型中直径为 6.4mm的冷、热区 ,且均匀区域未见伪影。给药 5 0min后开始采集所得患者图像清晰 ,但有时病灶难以与正常组织区分 ;给药 90min后开始采集所得患者图像清晰 ,对比度良好 ,能进一步明确病灶轮廓与数量。低活性较长采集时间也能获得较好的图像质量。结论 通过对活性和显像时间的平衡 ,明显降低了二者对PET/CT图像质量的影响。给药与开始采集数据的时间间隔最好在 90min以上。Objective To probe the image quality of PET/CT effected by injected activity and scanning time from phantom and patients' studies. Methods For phantom studies, a Jaszczak phantom with smaller diameter of cold and hot inserts were imaged in 3 different scanning conditions. The first was higher activity level and shorter scanning time (81.918 MBq、2 min per bed), the second was recommended activity level and scanning time (42.654 MBq, 3 min per bed)by the manufacturer, and the last one was lower activity level and longer scanning time (15.947 MBq,4 min per bed). For patients' studies,the same patient position and scanning parameters were used after 50 min and 90 min with 18 F FDG injected, respectively. For both phantom and patients' studies,the corresponding X ray CT data were used for the attenuation correction,and the clinical whole body reconstruction modes were used for the relevant scanning data. Then the image quality, especially the spatial resolution and contrast of the images, were analyzed and compared visually and quantitatively. A volunteer with 74 kg was imaged with the total injected activity of 152.026 MBq and the scanning time of 3 min 30 s per bed. Results Diameter with 6.4 mm of cold and hot inserts were resolved in the 3 imaging methods with phantom studies, respectively.There were also no artifacts found in the uniformity regions. For patient studies,better image quality was contributed by the images acquired at 90 min after 18 F FDG injected,comparing with that of 50 min. Outline of part of the lesions in the 50 min's images was difficult to distinguish from the normal tissue, while the number and shape of lesions were increased and improved in the same area as the 90 min's images. The volunteer's images were also good enough for clinical diagnosis. Conclusion It was important to find out the tradeoffs between the total injected activity and scanning time to decrease the impact to the image quality of PET/CT. It was also a good idea to increase the interval more
关 键 词:体层摄影术 发射型计算机 图像质量 质量保证 质量控制 脱氧葡萄糖
分 类 号:R814.42[医药卫生—影像医学与核医学] R817.4[医药卫生—放射医学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.117