两种尿酸生物传感器的研制和比较  

Study and Comparison of Two Biosensor for Uric Acid

在线阅读下载全文

作  者:张旸[1] 禹萍[1] 许春向[1] 

机构地区:[1]哈尔滨医科大学公卫学院,哈尔滨150001

出  处:《传感技术学报》1994年第4期28-33,共6页Chinese Journal of Sensors and Actuators

摘  要:将大白鼠肝脏切片和尿酸酶分别固定在不同氧电极上,制成尿酸组织电极和酶电极,研究了它们的测试条件、分析性能和应用情况.结果:组织电极的最佳测试条件为33℃,pH=9,Tris-HCl缓冲溶液,线性范围3.33~56.67μg/mL,K_M=3.3×10^(-4)mol/L;酶电极的最佳测试条件为28℃,pH=8.8,Na_2B_4O_7-(NH_4)_2SO_4缓冲溶液,线性范围6.67~133.3μg/mL,K_M=2.8×10^(-4)mol/L.二种电极测量值的变异系数均小于4%,电极性能各有优缺点.用它们测定质量控制血清,测量值均在示值范围内,有望应用于临床.An enzymic electrode and a tissue electrode were made by mounting uricaseand big mouse fresh liver onto the different oxygen e1ectrodes. The testing conditions,analytic properties and app1ications were measured and studied. The optilnal conditionsof the tissue electrode and that of enzymic electrode were respectively determined bel-low:33℃, pH= 9(Tris-HCl buffer); linear range 3.3-56.67μg/mL, 28℃, pH = 8. 8(Na2B4O7-(NH4)2) SO4 buffer), linear range 6.67-133.3μg/mL. The CV of detectingvalues of the two electrodes all are less than 4% and the Michaelis constants of the en-zymic e1ectrode and tissue electrode were 2. 8×10^(-4) mol/L and 3. 3 × 10^(-4) mol/L respec-tively. The data of the measurments in detecting control serum with the two electrodeswere all in allowing range. With further study and development the two electrodes canbe used in clinic.

关 键 词:尿酸 生物传感器 传感器 

分 类 号:TP212[自动化与计算机技术—检测技术与自动化装置]

 