EDTA-K_2抗凝剂对PLT检测结果的影响  被引量:2

The effect of EDTA-K_2 on the PLT counting

在线阅读下载全文

作  者:徐立山[1] 何虎[1] 赵明[1] 王立春[2] 

机构地区:[1]中国医科大学附属第一医院输血科,110001 [2]沈阳医学院奉天医院检验科

出  处:《辽宁医学杂志》2005年第2期62-63,共2页Medical Journal of Liaoning

摘  要:目的探讨EDTAK2抗凝剂对PLT读数的影响。方法用CD1700全自动血细胞分析仪检测EDTAK2抗凝静脉血的PLT结果与显微镜镜检不符个别患者,EDTAK2抗凝静脉血的PLT同时动用CD1700全自动血细胞分析仪预稀释功能及人工计数末梢血的PLT。结果用t检验方法得出EDTAK2组与预稀释组、人工计数组之间有极其显著性差异(P<0.01),而预稀释组、人工计数组之间则无显著性差异(P>0.05)。结论EDTAK2抗凝剂致偶见患者PLT假性降低。应用直接取末梢血,进行人工计数或动用CD1700预稀释功能,确保结果的准确性。而动用CD1700预稀释功能更方便、快速。Objective To investigate the effect of EDTA-K_2 on the PLT counting. Methods Count the PLT number of venous blood (with anticoagulant: EDTA-K_2 ) using CD-1700 automatic blood analyzer with its pre-dilution function, and with manual counting .Results T test showed that the counting between the First and the last two has significant difference (P<0.01). and the counting between the last two has no significant difference (P>0.05). Conclusion EDTA-K_2 can occasionally cause False decrease in the PLT counting. in order to get more accurate result , manual counting or pre-dilation function is necessary. However, pre-dilution function is more convenient and Fast.

关 键 词:抗凝剂 CD-1700全自动血细胞分析仪 检测结果 EDTA-K2抗凝静脉血 预稀释功能 PLT假性降低 显著性差异 人工计数 显微镜镜检 分析仪检测 t检验方法 末梢血 准确性 患者 数组 

分 类 号:R446.11[医药卫生—诊断学] R446.113[医药卫生—临床医学]

 