放置含铜硅胶V型宫内节育器的远期临床观察  被引量:7

Long term clinical observation on VCu200 in-trauterine devices

在线阅读下载全文

作  者:吴学浙[1] 陆健[1] 李慧沁 

机构地区:[1]上海市第一妇婴保健院

出  处:《中华妇产科杂志》1994年第9期545-547,共3页Chinese Journal of Obstetrics and Gynecology

摘  要:对放置国产含铜硅胶V型宫内节育器(IUD)5~18年的妇女350例(VCu200IUD组),与放置同样年限不锈钢金属IUD且同期取器的妇女350例(金属IUD组)的子宫内膜变化及IUD形态进行对照研究。结果:VCu200IUD组的妊娠率明显低于金属IUD组,虽然部分VCu200IUD出现铜丝氧化、融蚀,中心扣断脱致IUD变形及硅胶管钙盐沉积,但放置VCu200IUD妇女的子宫内膜无恶性病变发生。提示:如经x线检查IUD形态及铜丝尚较完整,且放置IUD的妇女无月经紊乱发生,仍可继续放置VCu200IUD。comparative study was carried out in 350 womenusing VCu200 intrauterine devices (IUD) and 350women using stainless steel ring (SSR) IUD for 5~18years. The endometrial changes and shapes of IUD werestudied. The results showed that the pregnancy ratewas significantly lower in VCu200 group after 5 years ofinsertion. In spite of the oxidation , errosion or breakageof the copper wire , calcium deposit of even deformationof the IUD, they were not associated with an increasedincidence of endometrial carcinoma and pelvic inflamma-tory disease. It is suggested that VCu200 IUD need notto be removed if the IUD keeps normal shape and thecopper wire remains intact by X ray examination.

关 键 词:宫内避孕器 病理学  

分 类 号:R169.41[医药卫生—公共卫生与预防医学]

 