离子色谱法快速测定饮水中毒副产物亚氯酸盐、氯酸盐和溴酸盐  被引量:12

Determination of chlorite, chlorate and bromate of byproduct disinfection in drinking water by ion chromatography

在线阅读下载全文

作  者:颜金良[1] 施家威[1] 周凯[1] 颜勇卿[2] 

机构地区:[1]浙江省宁波市疾病预防控制中心,浙江宁波315010 [2]北京大学基础医学院,北京100083

出  处:《中国卫生检验杂志》2005年第6期675-676,共2页Chinese Journal of Health Laboratory Technology

摘  要:目的:本文采用梯渡淋洗离子色谱法分别测定饮用水中的亚氯酸盐、氯酸盐和溴酸盐。方法:选用IonpacAS11阴离子分析柱,0~12mmol/LKOH淋洗液,流速为1.0ml/min,电导检测器。结果:本法相关性好(r>0.9990),精密度高(RSD%<5.0),样品加标回收率为88.3%~102%,最低检出量0.008~0.010μg/ml。结论:该法操作简单、快速、准确、灵敏、适用于大批量水样分析。Objective:To develop a mehtod for simultaneous determination of chlorite, chlorate and bromate in drinking water.Methods:The method was based on IC separation under the condition of isocratic eluent with 0~12 mmol/L KOH and flow rate of 1.0 ml/min by using the IonPAC AS11 analytical column and suppressed conductivity detector.Results:This method had good relativity (r>0. 9990) and good analytic precision (RSD%<5.0). The samples′average rates of recovery were 88.3%~102% and the detection limit was 0. 008~0.010 μg/ml.Conclusion:Its advantages are simple operation, quickness, high sensitivity and precise results.This method can be applied widely and suitable for trace analysis (μg/L) of water samples.

关 键 词:离子色谱法 亚氯酸盐 溴酸盐 毒副产物 快速测定 饮水 电导检测器 加标回收率 最低检出量 离子分析 水样分析 饮用水 min 相关性 精密度 大批量 

分 类 号:R123.1[医药卫生—环境卫生学] R123.6[医药卫生—公共卫生与预防医学]

 