摘 要:为了消除偶氮胂Ⅲ在酸性条件下测定血清钙所引起的蛋白质沉淀,报道一种在碱性条件下用偶氮脾Ⅲ测定血清钙的方法。在碱性条件下钙与偶氮胂Ⅲ的络合物在650nm波长的摩尔消光系数大大增加。加入8-羟基喹啉-5-磺酸可避免镁的干扰。在血清胆红素高达500μLmol/L、血红蛋白12g/L及脂肪乳剂Intralipid高达10g/L时,均对本法无显著干扰。钙标准曲线在0~5.0mmol/L范围内线性良好。该法与邻甲酚酞络合酮测定法所得结果相关性好。We propose a method for determination of calcium in serum using Arsenazo Ⅲ in alkaline medium. In the alkaline medium,the molar absorptivity of the Arsenazo Ⅲ complex with calcium at 650nm in-creases significantly. The interference from magnesium is avoided by using the 8-HQS. The assay showed no interference from as much as 5 00μmol/L of bilirubin,12g of haemoglobin and 10g of in-tralipid per liter. The standard curve for calcium was linear from 0 to 5.0mmol/L. The results from proposed method showed good agreement with results by OCPC procedure.