流动注射停流荧光动力学分析法测定痕量铁的研究  被引量:3

Studies on Flow-injection Stopped-flow Florescence Kinetic Method for Determination of Trace Iron

在线阅读下载全文

作  者:黄厚评[1] 蔡汝秀[1] 曾云鹗[1] 高登云[1] 

机构地区:[1]武汉大学化学系

出  处:《高等学校化学学报》1995年第1期47-49,共3页Chemical Journal of Chinese Universities

基  金:国家自然科学基金

摘  要:利用自行组装的流动注射停流荧光动力学分析装置,研究了铁(Ⅲ)与3,5-二溴水杨基荧光酮的荧光动力学反应性质,提出了一种快速、简便、精确地测定痕量铁的新方法.其线性范围为0.020~0.250μg/mL,线性相关系数为0.999,检测限为10ng/mL,以0.200μg/mL的铁(Ⅲ)作精度试验,R.S.D%=2.1%(n=11).该方法已成功地用于铝合金及桃树叶中痕量铁的测定,结果分别与ICP法和标准参考结果相符.A new flow-injection stopped-flow fluorescence kinetic method for determination of trace iron(Ⅲ) was proposed. The method is based on monitoring the binary fluorescence complex formation of Fe(Ⅲ) with 3,5-dibro mosalicylfluorone(DBSAF) at 510 nm/510 nm in an acidic medium.The linearity range is between 0.020 and 0.250 μg/mL with a correction coefficient of 0.999 and a detection limitation of 10 ng/mL. The method has been ap plied to the determination of Fe(Ⅲ)in alluminum alloy and peach leaves, the results are in agreement with ICP method and standard reference value, respectively.

关 键 词:荧光分析  二溴 水杨基荧光酮 

分 类 号:O614.811[理学—无机化学] O657.31[理学—化学]

 