检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]燕山大学电气工程学院,河北秦皇岛066004 [2]大庆石油学院分院,河北秦皇岛066004
出 处:《应用光学》2005年第5期10-12,共3页Journal of Applied Optics
基 金:国家自然科学基金(60272027);黑龙江省自然科学基金资助项目(F0312)
摘 要:根据西维因农药受到紫外光激发能够发出荧光的特征,利用脉冲氙灯作为激发光源,利用光纤进行传输和探测荧光,并结合电荷耦合器件(CCD)光谱探测技术,设计出了一套能够测量农药含量的荧光光谱仪器.分别利用稳态光谱仪和该仪器实现了对黄瓜中西维因农药残留的快速测定.实验结果表明,在激发波长为319 nm和荧光发射波长为647 nm情况下,西维因的线性范围为0.0~120 μg/L,最低检出限LOD为5×10-7 μg/L,线性相关系数r为0.9991(S/N=5).该仪器可满足荧光检测的需要.A fluorescence spectrometer which can measure pesticide concentration is designed by using a xenon lamp as excitation light-source,employing optical fibers to transmit and detect fluorescence and combining with CCD spectrumdetermination technology based on fluorescent characteristics of carbaryl that can emit fluorescence as it is excited by UV rays. Moreover,the rapid measurement of carbaryl residues in cucumbers is implemented with a steady-state fluorescent spectrograph and the new system respectively. The results indicate that the linear concentration range is 0.0-120.0 μg/L for carbaryl, the detection limit (LOD) is 5× 10^-7 μg/L,the linear correlation coefficient r is 0. 9991 as the excitation wavelength is 319 nm and the fluorescence emission wavelength is 647 nm. The data prove up to the hilt that the instrument meets the demand of fluorescence detection.