流动注射-在线氧化-电感耦合等离子体原子发射光谱法测定碘  被引量:14

Determination of Iodide by Flow Injection Analysis on Line Oxidation-Inductively Coupled Plasma-Atomic Emission Spectrometry

在线阅读下载全文

作  者:王化南[1] 尚福全 

机构地区:[1]哈尔滨工业大学分析测试中心,哈尔滨150001

出  处:《分析化学》1996年第3期288-291,共4页Chinese Journal of Analytical Chemistry

摘  要:本文把流动注射与在线氧化生成挥发性碘技术结合起来,不仅可以提高测定的灵敏度,而且可以减少氯化钠对测定的干扰。检出限达0.04mg/L。本法简便快速,易于操作,灵敏度高。The techniques of flow injection analysis and on-line oxidation of iodide were con bined and studied in this paper. A new analytical method of determining iodine was estab lished, in which not only the sensitivity can be enhanced a hundred times , but also the ad sorption of iodine molecules on the system and the interferences of NaCl were eliminited.The detection limit is 0. 04 mg/L. The method is simple, fast and easy to operate.

关 键 词:流动注射  原子发射光谱 ICP-AES 

分 类 号:O613.44[理学—无机化学] O657.31[理学—化学]

 