调强放射治疗IMRT计划临床剂量验证298例  被引量:1

Clinical Dose Verification for IMRT Plans of 298 Cases

在线阅读下载全文

作  者:黎杰[1] 窦绍宇[1] 肖明勇[1] 

机构地区:[1]四川省肿瘤医院放疗中心,四川成都610041

出  处:《四川肿瘤防治》2005年第4期257-258,共2页Sichuan Journal of Cancer Control

摘  要:目的:探讨调强放射治疗(IMRT)剂量验证方法的改进方向。方法:298例IMRT计划(NOMOSPeacock系统)靶区剂量计算值与验证实测值比较,得出相对误差结果。结果:298例验证计划中91.3%在平均误差±1.63%范围内,与实测剂量相符合。讨论:验证计划中电离室位置设置、测量模体、电离室体积的选用等,都需要进一步探讨和改进。Objective: To discuss investigate for improvement of IMBT plan dose verification. Methods: The dose of IMRT plans (NOMOS Peacock system) was compare with dose measurement and the relative error was obtained. Results: Average error of 91.3% verification plans is in range of 1.63%, which accorded with the results of dose measurement. Discussion: In verification of plans, location of ionization chambers, phantom and volume of chamber all need mere study and improvement.

关 键 词:验证计划 IMRT 误差 