躯干等速肌力测试重测信度研究  被引量:14

A study on the reliability of isokinetic test of the trunk muscles

在线阅读下载全文

作  者:成鹏[1] 黄杰[2] 仇瑶琴[1] 郎海涛[1] 

机构地区:[1]第二军医大学附属长海医院康复医学科,200433 [2]湖北省宜昌市第一人民医院康复医学科

出  处:《中国康复医学杂志》2006年第1期50-52,共3页Chinese Journal of Rehabilitation Medicine

摘  要:目的研究躯干肌等速测试的信度并初步分析影响测试的有关因素,为进一步深入研究奠定基础。方法20例自愿受试者经同一检测者利用同一程序、同一检测方法在一周内分别测试其躯干的等速力学指标,使用组内相关系数对结果进行信度分析。结果60°/s和180°/s收缩时重复测试信度满意,240°/s收缩时伸肌重复测试信度满意,但屈肌稍差。结论等速测力装置用于躯干的力学特征评定有较好的重复信度,尤其应选择低速和中速收缩,少用高速收缩。Objective: To study the reliability of isokinetic test of the trunk muscles, and to analyze the factors which may influence the isokinetic trunk test.Method:20 volunteers were tested twice by a same person with a same procedure, a same method within a same week. ICC was analyzed about their PT. PT/BW, TW, TW/BW and AP. Result: The reliability of test-retest of 60°/s and 180°/s was satisfied, and the reliability of extension of 240°/s was also satisfied but the flexion was inaccurate. Conclusion: The isokinetic test of the trunk muscles have good repeating reliability. The low and medium velocity test should be chosen,while high velocity test may be inaccurate.

关 键 词:等速测试 躯干 信度 

分 类 号:R49[医药卫生—康复医学]

 