检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]深圳市肝病研究所,深圳518020
出 处:《中国医疗器械信息》2006年第2期40-40,49,共2页China Medical Device Information
摘 要:目的:比较三种荧光定量PCR检测仪的偏倚。方法:采用实时荧光定量PCR,随机选择我院门诊和住院病人标本42例,用同一种试剂抽提DNA摸板后,分别在PE7000、伯乐及罗氏定量PCR扩增仪平行测定。结果42份样品在三种PCR检测仪上所测定HBVDNA拷贝/ml对数均值分别为5.625±1.449、5.420±1.960、5.803±1.536,三组之间比较无差异(P>0.1)。以测定拷贝数值≤或≥5.0×105/ml将其分为二组,在标本HBVDNA含量≤5.0×105拷贝/ml,伯乐测定值偏低,与罗氏比较差异有显著性意义((P<0.05))。结论伯乐仪器对HBVDNA低拷贝含量标本检测值偏低,PE7000型仪器有操作简单、耗材低廉、检测数据稳定及价格适中等优点,更适合于临床。Objective: To evaluate PE700/ABI and icycler/BioRad and Lightcycler/Roche. three fluorescent quantitative machines. Methods: Specimens were analyzers with one reaget by the three fluorescent quantitative machiines. Results: all the results obtained from the three machines were not significant. But 23 patients with the HBV-DNA value less than 5^* 10^5 copies/ml were significant difference between icycler/BioRad and Lightcycler/Roche .conclusion among of them PE7000 is much better.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.222