石墨炉原子吸收测定生物样品中痕量钴的基体改进剂的比较  被引量:3

Graphite furnace atomic absorption spectrophotometry for determination of trace cobalt in biological sample and matrix modifier

在线阅读下载全文

作  者:高振宗[1] 李玉琴[1] 

机构地区:[1]陕西师范大学实验中心

出  处:《陕西师范大学学报(自然科学版)》1996年第3期57-59,共3页Journal of Shaanxi Normal University:Natural Science Edition

基  金:陕西省卫生厅科研资助

摘  要:在比较钴的基体改进剂的基础上,选定了NH4VO3为基体改进剂,拟定了用石墨炉原子吸收法测定人发中痕量钴的分析方法.结果表明,方法的线性范围为0~320ng/mLCo2+,其特征量为1.3×10-3μg/(mL·1%).In determination of trace cobalt with graphite furnace atomic absorption spectrophotometry(GFAAS), the effect of matrix modifier of some reagents has been studied and compared. NH 4VO 3 is selected as the matrix modifier for changing the atomization process of sample of the biology. The sensitivity of the method is raised. The sensitivity is 1.3×10 -3 μg/(mL·1%). The calibration curve is linear for concentration 0~320 ng/mL Co 2+ . The relative error is ±2.7%. for the determination of trace Co in standard biological samples.

关 键 词:石墨炉原子吸收 生物样品 测定 基体改进剂 

分 类 号:Q581[生物学—生物化学] Q-331

 