两种血清唾液酸测定方法的比较  被引量:2

The compare of results between two tests of serum sialic acid.

在线阅读下载全文

作  者:宁婷[1] 殷路[2] 杜敏[1] 殷学军[3] 

机构地区:[1]新疆生产建设兵团总医院,830002 [2]武警新疆总队第三支队卫生队,843000 [3]兰州军区乌鲁木齐总医院,830000

出  处:《中国优生与遗传杂志》2006年第5期26-27,30,共3页Chinese Journal of Birth Health & Heredity

摘  要:本文采用两种唾液酸测定方法,比较了正常人群体(386例)血清唾液酸(SA)的吸光度、含量。结果表明,方法一测定的平均吸光度为A=0.15±0.028(相应的SA含量为1.31±0.251mmol/L),位于光度计准确区域以外;方法二测定的平均吸光度A=0.24±0.030(相应的SA含量为2.13±0.272mmol/L),位于光度计准确区域之内。两种方法测定386例正常人血清平均吸光度间、平均含量间有非常显著的统计学差异(P<0.01)。特别是在血清唾液酸增高的肿瘤患者血清测定时,方法二优于方法一。Objective: To compare the absorptivities (A) of 386 normal human serumwith two different tests, contents too. Results : ( 1 ) The average A was 0. 15 ± 0. 028 with method I ( corresponding content was 1.31 ± 0. 251 mmol/L), it is out of the standard test range (0. 2 ~0. 85) of Photometer; compare to this, A was 0. 24 ±0. 030 ( corresponding content was SA =2. 13 ± 0. 272mmoL/L) with method Ⅱ wh/ch is in the required range, the average absorptivity of serum from 386 normal human serum through two methods revealed clear difference ( P 〈 0. 01 ) . (2) The results from method Ⅱ were more stable and relable than method Ⅰ when test serum from tumor patients.

关 键 词:血清 唾液酸 吸光度 含量 

分 类 号:R446.11[医药卫生—诊断学]

 