基于NTN校准技术的脉冲标准法  被引量:2

The Pulse Standard Method Based on "Nose-to-Nose"Calibration Procedure

在线阅读下载全文

作  者:王跃佟[1] 刘明亮[1] 高小珣[2] 郁月华[2] 缪京元[2] 朱江淼[1] 

机构地区:[1]北京工业大学电子信息与控制工程学院,北京100022 [2]中国计量科学研究院,北京100013

出  处:《计量学报》2006年第3期266-268,共3页Acta Metrologica Sinica

摘  要:脉冲标准法是校准宽带取样示波器的一种行之有效的方法,但它需要波形参数已知的脉冲信号源,当取样示波器的带宽很宽时,这种脉冲信号源是很难得到的。近年来发展起来的NTN校准技术可以不受带宽的限制,但是它只限于对某些结构的取样示波器的校准。提出了一种基于NTN校准技术的脉冲标准法,该法克服了上述限制,可以实现更多种类的宽带取样示波器的校准。实验结果表明,该方法是非常有效的。The "pulse standard" method is an effective method on broad-hand sampling oscilloscopes calibration, but a pulse generator with an accurately known pulse shape is needed in this method. With the development of broad-hand sampling oscilloscopes, the standard pulse generator is very difficult to obtain. In recent years a method called "Nose-to-Nose" calibration procedure is not limited by the bandwidth, but it is only suitable for some of the sampling oscilloscopes calibration. A new kind of "pulse standard" method which based on the "Nose-to-Nose" calibration procedure is proposed. This new method which is suitable for more kind of broad-hand sampling oscilloscopes calibration and not limited by the standard puLse generator. The experiment result indicates that the method is extremely effective.

关 键 词:计量学 NTN校准技术 取样示波器 幅频响应 带宽 

分 类 号:TB973[一般工业技术—计量学]

 