摘 要:用ICS-Ⅱ型手动方式,采用速率比浊原理,检测AT-Ⅲ含量(试剂自配),是对进口仪器的开发利用,也是目前检测AT-Ⅲ最经济、最快速、最简易的方法。我们用本方法与火箭电泳法同时测34例,结果高度相关r=0.97。本法精密度试验,说明试剂稳定性极好,平均回收率为96.7%,30例健康人参考值为304±62(x±s)。用本法测定81例结果的mg/dL浓度和RU值做直线回归,r=0.91,回归标准差s=13.9,回归变异系数Sb=7.8,Y=54.4+5.4X。本法快速、准确,用血标本及抗血清量极微,节约人力、物力,方便临床,是目前测AT-Ⅲ较理想的方法。Determination of AT-Ⅲ content by manual rapid model ICS-Ⅱ and speed rate heterometry principle is the development and utilization of import instruments, also a economic, rapid and convenient method.We tested 34 cases by the method and the rocket electrophoresis.The result showed that the two ways were highly relative ( r =0 98), the fineness assay of the method demonstrated that the stability of the reagent was excellent, the average recover rate was 96 7 per cent.The reference value of 30 healthy individuals was 304±62( ±s ).The concentration (mg/dl) and RU value of 86 cases determined by this method were analysed statistically with linear regression, we got r =0 99, regression standard deviation S=13 9, regression variation coefficient sb=7 8, y=54 4±5 4x.This way is rapid and precise, needs a little blood sample and anti serum, saves labour power and material resources, makes things convenient for the clinic, and also is an ideal method for testing AT-Ⅲ at present time.