检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:郑洪涛[1] 汤志勇[1] 李威[1] 王迪民[2]
机构地区:[1]中国地质大学材料科学与化学工程学院,武汉430074 [2]宜昌地质矿产研究所,宜昌443003
出 处:《理化检验(化学分册)》2006年第9期759-761,共3页Physical Testing and Chemical Analysis(Part B:Chemical Analysis)
摘 要:研究了一种以NH4F为掩蔽剂掩蔽Fe(Ⅲ),用732型阳离子交换树脂(CSR)吸附Fe(Ⅱ),从而分离测定Fe(Ⅲ)与Fe(Ⅱ)的试验体系。设计了一套新颖的用于该体系的全自动流动注射流路,实现了Fe(Ⅲ)与Fe(Ⅱ)的连续测定。Fe(Ⅲ)和Fe(Ⅱ)的相对标准偏差(n=7)分别为3.37%和1.95%,检出限分别为0.07和0.01 mg.L-1,分析速度为25个样/h。A method of flow injection ion exchange separation followed by flame atomic absorption spectrometry for the speciation analysis of iron was studied in this paper. Iron( Ⅲ ) in sample solution was masked by addition of NH4 F and Fe( Ⅱ ) was adsorbed on the micro column packed with 732^# cation exchange resin. The adsorbed Fe( Ⅱ ) was then eluted with 6 mol . L^-1 HC1 solution. Fe( Ⅲ) and Fe( Ⅱ ) were successively determined by flame atomic absorption spectrometry. Detection limits of this method were found as 0.07 mg . L ^-1 for Fe( Ⅲ ) and 0.01 mg. L^-1 for Fe( Ⅱ ), and RSD's (n= 7) of 3. 37% for Fe( Ⅲ ) and 1. 95% for Fe( Ⅱ ). Using the proposed method, about 5 samples can be analyzed in an hour.
正在载入数据...