ADVIA2120全自动血细胞分析仪测定网织红细胞的评价  被引量:1

Evaluation of performance of ADVIA2120 blood-counter system for detection of reticulocytes.

在线阅读下载全文

作  者:许睿[1] 廖春盛[1] 

机构地区:[1]广东省中医院珠海医院,广东珠海519015

出  处:《中国热带医学》2006年第9期1730-1731,1733,共3页China Tropical Medicine

摘  要:目的 对ADVIA2120全自动血细胞分析仪测定网织红细胞的性能进行评价。方法用ADVIA2120全自动血细胞分析仪测定高、中、低值网织红细胞样本,进行重复性、携带污染率及稳定性测试,并与手工计数结果比较。取高值样本在技术指标范围内按比例稀释后测定各浓度,进行线性评价。结果 重复性:低值指标高核酸浓度网织红细胞比率CV值较大(13.82%~17.4%),其它各项指标的CV均〈8.19%;线性r=0.9989;稳定性显示,4℃低温24h内网织红细胞百分率计数CV为3.8%;携带污染率为0.22%;与手工法百分率结果比较,相关性好(r=0.9962),差异无统计学意义(P〉0.05)。结论 ADVIA2120全自动血细胞分析仪测定网织红细胞携带污染率低、重复性、稳定性、线性好,能满足临床需要,可替代手工计数,但不能区分疟原虫感染引起的干扰。Objective To evaluate the performance of ADVIA2120 for the test of reticulocytes. Methods ADVIA2120 was used for detection of high/normal/low reticulocyte samples and the reproducibility , the rate of carrying contamination and stability were observed. The results were compared with manual method. Results Reproducibility : except that the CV of high nucleic acid was 13.82 % - 17.4 %, the CV of the other indexes was less than 8.19 % ; linearty : r = 0.9979 ; stability : percentage CV of Ret at 4℃ in 24h was 3.8% ; the rate of carrying contamination was 0.22% ; compared with the manual methods : r= 0.9962, P 〉 0.05. Conclusion The state of carrying contamination by using ADVIA2120 blood - counter system in test of reticulocytes is low, stable and showing a nice linear correlation suitable for clinical use.

关 键 词:网织红细胞 全自动血细胞分析仪 疟原虫 

分 类 号:R446.11[医药卫生—诊断学]

 