检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
出 处:《中华放射医学与防护杂志》2006年第5期517-519,525,共4页Chinese Journal of Radiological Medicine and Protection
摘 要:目的估算使用前列腺增生治疗器人员(医生)职业照射的个人剂量上限和评价使用前列腺增生治疗器的辐射安全。方法用辐射防护仪表测量前列腺增生治疗器周围的辐射剂量[定向剂量当量H′(d)和周围剂量当量H*(d)],调查医院前列腺增生治疗器临床试用情况和测量临床试用中的辐射剂量来估算医生的个人剂量上限。结果患者治疗1疗程,医生个人剂量上限分别为6·0μSv(尿道型)和4·5μSv(直肠型)。结论在正常条件下使用前列腺增生治疗器的辐射安全是有保障的。Objective To evaluate the upper limit of individual dose to the people of occupational exposure(medical doctors) and appraise radiation safety on using prostate hyperplasia applicators. Methods The radiation dose ( directional and ambient dose equivalent ) around the source of the prostate hyperplasia had been determined with radiation protection instruments. According to inquired situation of clinical application of prostate hyperplasia applicators in hospitals and measured radiation dose in clinical probation period, the upper limit of individual dose of midical doctors had been evaluated as well. Results The upper limit of individual dose per person-course are 6.0 μSv for arethra-type and 4.5 μSv for rectum-type, respectively. Conclusions Under the normal conditions, the radiation safety can be ensured when operating the prostate hyperplasia applicators.
分 类 号:R144[医药卫生—公共卫生与预防医学]
正在载入数据...