应用于原子荧光光谱分析的空心阴极灯灯电流与激发光强度关系的实验  被引量:7

Experiments on Relationship between Lamp Current of Hollow Cathode Lamp and Exciting Light Intensity in Atomic Fluorescence Spectrometric Analysis

在线阅读下载全文

作  者:黄跃[1] 李可[1] 赵志华[1] 王莉坤[1] 魏建山[1] 

机构地区:[1]中国地质科学院地球物理地球化学勘查研究所,河北廊坊065000

出  处:《岩矿测试》2006年第4期384-386,共3页Rock and Mineral Analysis

摘  要:应用XGY-6080型全自动双通道原子荧光光谱仪,研究了As、B i、Sb、Hg、Se、Te、Pb、Sn、Ge、Cd和Zn等元素的高效特种空心阴极灯的一些性能特点,分析和讨论了各元素灯主、辅阴极电流与激发光强度之间的关系,为进行原子荧光光谱分析时空心阴极灯供电电流的选择提供一些参考依据。A model XGY-6080 double-channel automatic atomic fluorescence spectrometer was used to study the primary properties of hollow cathode lamps and systematically evaluate the effect of hollow cathode lamp current on the fluorescence signal intensity. The elements of As, Bi, Sb, Hg, Se, Te, Pb, Sn, Ge, Cd and Zn were involved in the experiments and the results from the experiments provided the guidance in choosing lamp current correctly in atomic fluorescence spectrometric analysis.

关 键 词:原子荧光光谱仪 高性能空心阴极灯 灯电流 荧光信号强度 

分 类 号:O657.31[理学—分析化学]

 