检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:李利荣[1] 吴宇峰[1] 杨家凤[1] 时庭锐 袁倩[1] 魏恩棋[1]
出 处:《中国环境监测》2007年第1期11-15,共5页Environmental Monitoring in China
基 金:科技部国家科技基础条件平台项目(2004DKA10150)
摘 要:建立了固相萃取、毛细柱气相色谱-电子捕获检测器定量分析水中硝基苯类化合物的分析方法。通过对固相萃取柱的选择、固相萃取条件(样品溶液的pH值、上样速度、上样体积、洗脱液选择及配比)的优化,得出了最佳实验条件。穿透体积可达1.5L,回收率在80%-105%,最小检出浓度在0.00037—0.072μg/L之间,均大大低于GB13194—91的规定。相对标准偏差在0.7%-5.3%之间。该方法准确、简单,重现性好、回收率高,污染小。A method for the determination of aromatic nitrocompounds in water has been established by solid-phase extraction(SPE), capillary gas chromatography with electron capture detector. Through the selection of SPE columns, the chosen of SPE conditions which include the pH value of water sample, load speed, load volume, elution solutions, etc, the optimum inspection conditions was got. The max load volumes can be 1.5L.The recoveries of every compounds was between 80% and 105%. The relative standard deviation ranged from 0.7% to 5.3%, the minimal detection concentration was between 0.00037 - 0.072μg/L, lower than those of GB13194- 91 's. The correlation is between 0.9993 - 0.9999.
分 类 号:X832[环境科学与工程—环境工程]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.222