胰岛素、C肽RIA检测中的室内质量控制  被引量:3

Intra-laboratory Quality Control of RIA for Insulin and C-Peptide

在线阅读下载全文

作  者:侯永生[1] 袁凤山[1] 李强[1] 

机构地区:[1]中国医科大学第一临床学院

出  处:《标记免疫分析与临床》1996年第1期33-36,共4页Labeled Immunoassays and Clinical Medicine

摘  要:本文介绍了胰岛素、C肽RIA检测中的三种室内质控方法,即质控血清监测法,两样本Youden图监测法和标准曲线参数监测法,并对其结果做了初步分析。三种室内质控方法的应用,保证了每一批测定结果的准确可靠、提高了不同批实验结果的可比性。本文还对胰岛素、C肽RIA检测中使用的试剂盒进行了方法学评价。胰岛素、C肽测定的批内CV分别为5.48%和9.98%;批间CV分别为10.54%和14.56%;平均回收率分别为102.6%和98.7%;灵敏度分别为1.25mU/L和0.02nmol/L。测定了184名正常人空腹血清胰岛素和C肽的含量,分别为8.35±4.47mU/L和0.25~1.32nmol/L。与文献报告的正常值接近。Three kinds of intra-laboratory quality control tests of RIA for serum insulin and C-peptide,i.e.monitoring by quality control serum,by two-sample Youden plotting and by standard curve parameter are described in the present paper.Analysis of the results demonstrated that three QC tests could ensure the precision of data within assay and improve the consistency between assays. Evaluation of insulin and C-peptide RIA kits showed the following results: insulin─intra-assay CV─5.48%.inter-assay CV 10.54%.recovery rate 102.6%,limit of detection 1.25mU/L and C-peptide,9.98%,14.56%,98.7%,0.02nmol/L,respectively.Serum level of insulin in 184 healthy subjects was 8.35±4.47mU/L and that of c-peptide 0.25 ̄1.32 nmol/L.

关 键 词:胰岛素 C肽 放射免疫分析 质量控制 糖尿病 

分 类 号:R587.104[医药卫生—内分泌] R446.6[医药卫生—内科学]

 