出  处:《医疗卫生装备》2007年第2期67-68,共2页Chinese Medical Equipment Journal

摘  要:由于本地区气候原因,设备PCB板时常会出现电位器变值和电路故障。开机出现黑屏,电源指示灯不亮,没有kV预置显示,没法选择条件,看不到Er1~Er7的故障代码。通过电路解析,分析设备的功能原理。以工作流程指示灯为主线,可快速判断和排除设备故障。Abstract PCB faults on X-ray machine occur frequently in our hospital due to humid climate, which lead to many malfunctions such as potentiometer failure, blank screen, lacking KV display, unavailable parameter choosing and disappearing of Erl -Er7 fault code. These malfunctions could be diagnosed and repaired quickly based on analyzing the status of the machine' s workflow indicator lights, as well as studying circuit and function principle of the equipment.