作 者:徐永清[1] 覃励明[1] 陆声[1] 赵卫东[2] 张美超[2] 朱跃良[1] 吴农欣[1]
机构地区:[1]成都军区昆明总医院全军骨科中心,昆明650032 [2]南方医科大学广东省医学生物力学重点实验室,广州510515
出 处:《中国临床解剖学杂志》2007年第3期332-334,共3页Chinese Journal of Clinical Anatomy
基 金:全军医药卫生科研基金(06MA140)
摘 要:目的:评价镍钛记忆合金腕骨四角融合器(NiTi shape memory alloy four-corner arthrodesis concentrator,NT-FCAC)的生物力学特性。方法:32侧头状骨、月骨、三角骨、钩骨标本分别采用NT-FCAC和克氏针固定,体外模拟四角融合术,858型双轴液压材料测试系统(858-MTS)测试固定强度;2件NT-FCAC施加20~200N循环拉伸载荷(1Hz),达到6%记录形变时的循环载荷次数;100N下压力和上提力加载NT-FCAC有限元模型,了解应力分布情况;分析试验结果,评价NT-FCAC生物力学特性。结果:NT-FCAC组固定强度大于克氏针组(P<0.05);2件NT-FCAC发生形变达到6%时,施加载荷次数分别为4239和4068次;应力分布集中在NT-FCAC体部与4个固定臂交界处。结论:NT-FCAC固定强度可满足四角融合术后不需外固定、早期功能锻炼的要求;疲劳度可满足四角融合后达到骨性愈合的需要;最易疲劳、断裂部位在NT-FCAC体部与4个固定臂交界处。Objective: To evaluate the features of NiTi shape memory alloy four-comer arthrodesis concentrator (NT-FCAC). Methods: 32 specimens of capitate, lunate, triquetrum and hamate were used in this investigation. Four-comer arthrodesis was carried out in each specimen in vitro, with 16 using NT-FCAC and 16 using K-wires. Their fixed strength was tested by a hydrolic pressure biomechanical system (858-MTS). Two NT-FCACs were fatigued with 20 -200N cycling pulling load (1Hz) delivered by 858-MTS and the deformations of them were carefully observed. The stresses were measured after the three dimensional finite elements of NT-FCAC were loaded with 100N. The biomechanical features of NT-FCAC were evaluated according to the test data. Results: The strengths were stronger in NT-FCAC group than in K-wires group (P〈0.05). When the two fixators (NT-FCAC) were deformed to 6% by pulling load, they had resisted the pulling load for 4 239 times and 4 068 times respectively. The junctures between the body and fixed-arms of NT-FCAC received the greatest stress. Conclusions: According to the biomechanical results, the early exercise can be carried out when the wrist is not need to be external fixed after four-comer arthrodesis with NT-FCAC. The four bones will be got to bony fusion under the range of fatigued strengths of NT-FCAC. It seems that the junctional zones between the body and fixed-arms of NT-FCAC are more likely to be broken.
分 类 号:R318.01[医药卫生—生物医学工程] R318.08[医药卫生—基础医学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.217