日立7600-020E生化分析仪电解质模块应用评价  被引量:4

Applicable Evaluation on ISE Part of Hitachi 7600-020 Biochemical Analyzer

在线阅读下载全文

作  者:罗浔阳[1] 张劼[1] 孙兵[1] 王菁菁[1] 徐志晔[1] 

机构地区:[1]南京大学医学院附属南京鼓楼医院检验科,江苏南京210008

出  处:《医疗设备信息》2007年第5期53-54,43,共3页Information of Medical Equipment

摘  要:目的评价日立7600-020自动生化仪电解质模块测定钾钠氯离子的精密度、线性和携带率。方法依据美国临床实验室标准化委员会(National Committee for Clinical Laboratory Standards,NCCLS)EP6-A、EP9-A文件方法,计算线性方程和相关系数,并以此为依据进行结果的判断。结果该仪器电解质模块测定精度为结论钾CV批内=0.68%、钠CV批间= 0.53%,氯CV批内=0.51%、钾CV批间=0.96%,钠CV批内=0.66%、氯CV批间=0.85%,且线性方程回归系数r均大于0.999。结论用日立7600-020电解质模块测定样本快速准确,具有测定线性范围广、携带率低的优点。Objective To evaluate the precision, linearity and sample carrying rate of ISE part of Hitachi 7600-020 biochemical analyzer. Method Based on the NCCLS EP6-A EP9-A document, the bias evaluation, correlation coefficient, linear regression equation were calculated. Results The CV value of the same items measured in different analytic system is (0.51-0.68)%, the CV of values between different days is (0.66-0.96)%. The correlation coefficients are more than 0. 999. Conclusion The instrument has the advantages of a good linearity and low sample carrying rate.

关 键 词:电解质模块 性能评价 生化分析仪 

分 类 号:TH776[机械工程—仪器科学与技术]

 