预真空蒸汽灭菌器B-D试验监测效果分析  被引量:1

Analysis of B-D test monitoring result on forevacuum steam sterilizer

在线阅读下载全文

作  者:张鲜利[1] 张红军[1] 付敏[1] 

机构地区:[1]沈阳军区总医院北陵临床部,沈阳市110031

出  处:《医疗卫生装备》2007年第6期64-64,共1页Chinese Medical Equipment Journal

摘  要:通过脉动预真空压力蒸汽灭菌器B-D试验监测,分析并找出影响B-D试验结果的不合格因素。采取有效方法,排除影响B-D监测结果因素,及时发现问题,对灭菌器进行检修。经常对预真空蒸汽灭菌器进行保养,严格管理,确保预真空压力蒸汽灭菌器灭菌按程序安全运行,保证灭菌质量合格,使临床使用安全、可靠。This paper analyzes the disqualification factors influencing B-D test results by using B-D test monitoring. Effective methods were taken to remove them and the problems were found in time for overhauling the sterilizer. The forevacuum steam sterilizer should be kept in good state and regulated strictly so that it can work in program with qualified sterilizing quality, thus ensuing a safe and reliable clinical application.

关 键 词:B-D试验 脉动预真空压力蒸汽灭菌器 监测 

分 类 号:R187[医药卫生—流行病学]

 