检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:曹磊[1] 张文艺[1] 刘柏群[1] 候长松[1] 赵兰才[1]
机构地区:[1]中国疾病预防控制中心辐射防护与核安全医学所辐射防护与建设项目评价室,北京100088
出 处:《中华放射医学与防护杂志》2007年第3期263-264,共2页Chinese Journal of Radiological Medicine and Protection
摘 要:目的探讨医用加速器机房中L型迷路和Z型迷路之间的差异。方法选择典型医用电子直线加速器作为研究对象,通过对加速器MV-X射线模式下工作条件的进一步分析,参考NCRP51号报告及有关文献中迷路设计方法进行研究。结果采用Z型迷路可以有效降低治疗室出入口的辐射水平。结论医用加速器机房采用Z型迷路设计是最优化设计的有益参考。Objective To discuss the differences on the design and evaluation between the L-type and Z-typo mazes of typical medical electric linear accelerator. Methods The study is conducted by choosing some typical medical electric linear accelerators used in China, further analyzing on the running conditions of the accelerators in the mode of MV-X-ray, and referring to the late NCRP Report 51 and other references. Results The radiation levels at the access to therapy room are effectively reduced by Z-type mazes. Conclusions The Ztype mazes are advisable during the optimization design.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.117