检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:张菊梅[1] 张希寒[1] 颜彦[1] 季安[1] 陈旻[1] 郭兴明[1]
出 处:《医疗卫生装备》2007年第7期29-30,共2页Chinese Medical Equipment Journal
基 金:重庆大学大学生创新基金资助项目;重庆大学生物工程学院创新人才培养基金项目资助
摘 要:对无创血压测量仪测量值误差异性展开分析。用与各种血压测量技术进行比较的方法,着重分析了振动法无创血压测量和柯氏音听诊法。使用振动法无创伤血压测量监护仪和柯氏音听诊法对样本人群进行了血压测量,分析所得数据,结果表明,基于振动法和柯氏音法无创血压测量仪测量值差异小,准确性也比较好。同时,也展望了无创血压装置未来的发展方向。This paper introduces and compares several blood pressure measurements, among which two noninvaslve blood pressure measurements based on oscillometric and auscuhatory methods are mainly compared. It also analyses the characteristics and advantages of oscillometric method as well as introduces a noninvasive blood pressure measurement monitor based on oscillometric. By using this monitor and a common auscultory method, we conducted a public survey. The result for statistical data analysis showed that the difference between oseillometric and auscuhatory methods are quite small, and the accuracy of them are both quite good .This paper also previews the future of the noninvasive blood pressure measurement