UF-100尿液分析仪在尿路感染诊断中的应用  被引量:11

Application of flow cytometer UF-100 in screening urinary tract infection

在线阅读下载全文

作  者:何发彬[1] 李阳[1] 阎衡[2] 项贵明[1] 

机构地区:[1]第三军医大学新桥医院检验科,重庆400037 [2]第三军医大学西南医院皮肤性病科

出  处:《检验医学与临床》2008年第2期76-77,共2页Laboratory Medicine and Clinic

摘  要:目的探讨UF-100尿液分析仪定量计数尿液中细菌在尿路感染诊断中的作用。方法将183份中段尿标本分别做尿细菌培养,同时用UF-100检测每份标本的细菌数,将两种方法得到的结果进行比较分析。结果以UF-100尿液分析仪细菌计数>6000/μl作为尿路感染的判断标准,与细菌培养比较,两者差异无统计学意义(P>0.05),其敏感性为73.68%,特异性为97.93%,阳性预期值为90.32%,阴性预期值为93.42%,准确度为92.94%。结论应用UF-100尿液分析仪对中段尿细菌数进行定量分析不失为尿路感染诊断一个快速、有效的方法。Objective To study the reliability of flow cytometer UF-100 in diagnosing urinary tract infection by counting the bacteria in urine. Methods 183 fresh midstream clean-catch urine samples were collected. Each sample was examined by bacterial cultures and flow cytometer UF-100 respectively. The results were compared with each other. Results When the bacteria of the UF-100 counting were 〉 6000μ1, there was no significant difference compared with the results of the bacterial culture(P〉0. 05). Its sensitivity and specificity were 73.68% and 97.93% respectively. Conclusion Using UF-100 to count the bacteria of the midstream urine is a rapid,effective method in diagnosing urinary tract infection.

关 键 词:尿分析 细菌培养 尿路感染 

分 类 号:R446.122[医药卫生—诊断学] R691.3[医药卫生—临床医学]

 