The check and error analysis of the BRDF experiment bench  被引量:2

The check and error analysis of the BRDF experiment bench

在线阅读下载全文

作  者:谢鸣 任虎 邹勇 李颖 谈和平 

机构地区:[1]School of Energy Science and Engineering,Harbin Institute of Technology

出  处:《Journal of Harbin Institute of Technology(New Series)》2007年第6期784-787,共4页哈尔滨工业大学学报(英文版)

摘  要:In this paper,a detailed introduction is given to the check method of the BRDF experiment bench which is built on our own. Measurement is made on the BRDF of the standard white plate made of polytetrafluoroethylene(PTFE)with reference to the existing standard white plate whose surface reflectance is known and by the method of theory approximate and relative comparison. On the basis of that,the BRDF value of the standard white plate in the wave band of 0.6328μm is given and the experiment bench is checked,the relative error of the experiment bench being within 20%.In this paper, a detailed introduction is given to the check method of the BRDF experiment bench which is built on our own. Measurement is made on the BRDF of the standard white plate made of polytetrafluoroethylene (PTFE)with reference to the existing standard white plate whose surface reflectance is known and by the method of theory approximate and relative comparison. On the basis of that, the BRDF value of the standard white plate in the wave band of 0. 6328μm is given and the experiment bench is checked, the relative error of the experiment bench being within 20%.

关 键 词:BRDF surface reflectance ERROR 

分 类 号:TU8[建筑科学]

 