应用哈希仪器消解测定水和废水中总磷  被引量:5

The Application of HACH Apparatus to the Total Phosphorus Determination in Water and Waste Water

在线阅读下载全文

作  者:王颖娜[1] 胡艳[1] 李德豪[1] 杨祥[2] 

机构地区:[1]茂名学院,广东茂名525000 [2]中国地质大学,湖北武汉430074

出  处:《茂名学院学报》2008年第1期14-16,共3页Journal of Maoming College

摘  要:建立了利用HACH公司的COD消解一比色管、DRB200 COD加热器及DR890便携式比色计测定水体总磷的方法。确定最佳的操作条件为:过硫酸钾质量分数4%,消解温度120℃,消解时间15min,抗坏血酸溶液的质量分数2%,钼酸盐使用液浓度为储备液1,4。该方法具有较好的精密度和准确度,检出限为0.039nag·L^-1,实际样品分析的相对标准偏差(RSD)在1.5%以下,加标回收率为90.0%.112%。This paper has brought about a method of determining the total phosphorus in water and waste water by using the COD digest - colorimetrie tube, DRB200 COD heater and DR890 portable colorimeter. The optimal operation condition was verified and the accuracy and precision was satisfactory: detection limit was 0.039 mg·L^- l ; relative standard deviation (RSD) of the specimen was below 1.5 % ; stan- dard recovery of this method was between 90.0% and 112%.

关 键 词:总磷 COD消解-比色管 DRB200 COD加热器 DR890便携式比色计 

分 类 号:O613.62[理学—无机化学]

 