机构地区:[1]Shang Institute of Applied Phsyics,Chinese Academy of Sciences,Shanghai 201800,China Graduate University of Chinese Academy of Sciences,Beijing 100049,China [2]Department of Engineering Physics,Tsinghua University,Beijing 100084,China [3]Shang Institute of Applied Phsyics,Chinese Academy of Sciences,Shanghai 201800,China
出 处:《Chinese Physics C》2008年第5期385-388,共4页中国物理C(英文版)
基 金:Development Plan of the State Key Fundamental Research(2002CB713600)
摘 要:RF cold test of a novel C-band cavity beam position monitor (PBM) to be used in the SDUVFEL Test Facility is described. The test results are presented and some characteristics discussed. The main parameters obtained are in reasonable agreement with the analytical estimations. Effective suppression of the common mode has been demonstrated. The position sensitivity over the test region of μ0.5 mm is about -21.58 dB/10 μm for the TM110 mode and is linear in the central region of the BPM cavity.RF cold test of a novel C-band cavity beam position monitor (PBM) to be used in the SDUVFEL Test Facility is described. The test results are presented and some characteristics discussed. The main parameters obtained are in reasonable agreement with the analytical estimations. Effective suppression of the common mode has been demonstrated. The position sensitivity over the test region of μ0.5 mm is about -21.58 dB/10 μm for the TM110 mode and is linear in the central region of the BPM cavity.
关 键 词:cavity BPM TM110 mode S-PARAMETER position sensitivity
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.171