全自动酶联免疫分析仪最佳操作模式探讨  被引量:2

Best Operation Mode of AP-960 Full-automated ELISA System

在线阅读下载全文

作  者:朱惠君[1] 徐志康[1] 刘和录[1] 杨小星[1] 刘镇平[1] 

机构地区:[1]深圳市沙井人民医院,广东深圳518104

出  处:《医疗卫生装备》2008年第5期99-100,共2页Chinese Medical Equipment Journal

摘  要:目的:选取AP-960全自动酶联免疫分析仪并对其上机操作模式进行探讨。方法:用加样开始模式(模式1)和加酶开始模式(模式2)进行上机检测,根据所消耗的配套吸嘴、需要的时间、实验结果以及结合常出现的警报进行对比。结果:2种模式所消耗的配套吸嘴分别为564个和84个;所需总时间分别为125 min和90 min。结论:选用加酶开始模式(模式2)比较适合本科室的实际情况。Objective To compare the operation modes of AP-960 full automated ELISA system, Methods Two models were tested: beginning with adding specimens (model 1) and beginning with adding enzymes (model 2). The following factors were compared: the consumption of suckers, usage of time, result of test and the usual alarm during the test. Results The consumptions of suckers were 564 and 84 respectively, The needing time of the former mode was 125 minutes and that of the latter mode was 90 minutes. Conclusion The mode beginning with adding enzymes(model 2) is better for the current status of our clinical laboratory.

关 键 词:AP-960 分析仪 操作模式 探讨 

分 类 号:TH776[机械工程—仪器科学与技术]

 