^20lTl定量门控心肌灌注体层显像与^99mTc-红细胞门控心血池显像测定左心室射血分数的对比研究  被引量:1

Comparison study of left ventricular ejection fraction measured by ^20lTl gated myocardial imaging and ^99mTc-red blood cell gated blood-pool imaging

在线阅读下载全文

作  者:袁建伟[1] 冯彦林[1] 贺小红[1] 余丰文[1] 刘德军[1] 温广华[1] 黄克敏[1] 

机构地区:[1]广东省佛山市第一人民医院核医学科,528000

出  处:《国际放射医学核医学杂志》2008年第2期86-88,共3页International Journal of Radiation Medicine and Nuclear Medicine

摘  要:目的探讨^20lTl定量门控心肌灌注体层显像与^99mTc-红细胞门控心血池显像测量左心室射血分数(LVEF)的相关性。方法72例受检者接受^20lTl静息门控心肌灌注体层显像,用AUTOQUANT 4.21软件测量LVEF,并与24h内的静息^99mTc-红细胞平衡法门控心血池显像结果进行比较。结果①门控心肌灌注体层显像与门控心血池显像测量LVEF值的结果呈明显正相关(r=0.554,P=0.000),两种方法无统计学差别(t=1.194,P〉0.05)。②不同疾病组之间两种测量方法无统计学差异(P值均大于0.05)。③门控心肌灌注体层显像及门控心血池显像测量的LVEF值分别为(64.68±10.77)%和(62.46±8.99)%,门控心肌灌注体层显像测量的LVEF值要比门控心血池显像高出3.55%。结论^20lTl控心肌灌注体层显像与^99mTc-红细胞门控心血池显像测量LVEF值的相关性好且结果准确,但门控心肌灌注体层显像的LVEF测量值要稍高于门控心血池显像。Objective To explore the correlation of automatic quantification of left ventricular ejection fraction (LVEF) from gated myocardial perfusion SPECT with ^20lTl and rest equilibrium gated blood-pool imaging with ^99mTc-red blood cell. Methods Rest gated myocardial perfusion SPECT with ^20lTl was performed on 72 cases. AUTOQUANT 4.21 software was used to measure LVEF. Rest equilibrium gated blood-pool imaging was also performed within 24 h. The values of LVEF measured by two methods were calculated and compared. Results ①There were strong correlation between the LVEF measured by rest gated myocardial perfusion SPECT and rest equilibrium gated blood-pool imaging(r=0.554, P=O.O00). There was no significant difference between the two methods (t =1.194, P〉0.05). ②There was no significant difference between the two methods among different disease groups. ③The values of LVEF measured by two methods were (64.68 ±10.77 )% and (62.46±8.99)%. The values of LVEF measured by gated myocardial perfusion SPECT was higher 3.55% than rest equilibrium gated blood-pool imaging. Conclusion There was good correlation between the LVEF measured with gated myocardial perfusion SPECT by AUTOQUANT 4.21 software and gated blood-pool imaging. The LVEF can be accurately evaluated using gated myocardial perfusion SPECT with ^20lTl. The value of LVEF measured by gated myocardial perfusion SPECT was higher than rest equilibrium gated blood-pool imaging.

关 键 词:门控血池显像 门控心肌灌注显像 铊放射性同位素 体层摄影术 发射型计算机 单光子 

分 类 号:R817.4[医药卫生—影像医学与核医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象