动态体温遥测装置与水银柱式体温计测温的对比研究  被引量:5

Comparative Study on the Temperature Measurement Methods by Using the Dynamic Distant Thermometer and Mercury Thermometer

在线阅读下载全文

作  者:郑文瑶[1] 孟庆义[1] 

机构地区:[1]中国人民解放军总医院急诊科,北京100853

出  处:《河北医学》2008年第7期796-798,共3页Hebei Medicine

摘  要:目的:了解动态体温遥测装置与水银柱式体温计体温测量结果有无差异。方法:用动态体温遥测装置与水银柱式体温计同时测量100名急诊留观病人腋窝温度,对结果进行相关性分析。结果:动态体温遥测装置组患者的平均温度为36.9±0.7°C,与水银柱式体温计测量结果(36.8±0.7°C)相近,两组测量数值间无统计学差异(P>0.05);且两组测量数值直线相关分析表明,呈高度相关(r=0.981,P<0.001)。结论:动态体温遥测装置与水银柱式体温计相比同样有效,并且使用更安全、简便。Objective: To determine whether there is any difference between the temperature reading measured by the dynamic distant thermometer and mercury thermometer. Method: The body temperatures in the 100 emergency patients were measured by the dynamic distant thermometer and mercury thermometer. The reading data were analyzed statistically. Results: The average body temperatures with the dynamic distant thermometer was 36.9 ±0.7℃ , which was no significant difference with the results measured by mercury thermometer (36.8 ±0.7℃, P 〉0.05 ). The body temperature measured by dynamic distant thermometer was correlated highly with the mercury thermometer ( r = 0.981, P 〈 0.001). Conclusion : The dynamic distant thermometer is an accurate method for measuring body temperature, and maybe safer and easy use than the mercury thermometer in the daily clinical practice.

关 键 词:体温计 体温 测量 

分 类 号:R472[医药卫生—护理学]

 