SPOTCHEM-4410分析仪在卫生运输船环境中的应用分析  

Application of the SPOTCHEM-4410 analysator in medical transport ship environment

在线阅读下载全文

作  者:唐大海[1] 吴玲[1] 贺嘉蕾[1] 

机构地区:[1]解放军第四一一医院检验科,上海200081

出  处:《中华航海医学与高气压医学杂志》2008年第3期172-173,共2页Chinese Journal of Nautical Medicine and Hyperbaric Medicine

摘  要:目的评价SPOTCHEM-4410分析仪在舰船环境中的临床分析性能。方法通过SPOTCHEM-4410生化仪在医院实验室内和在航行中检测的结果分别进行比对,评价其结果准确度、批内精密度和批间精密度。结果SPOTCHEM-4410分析仪在医院实验室内和在航行舰船环境中检测总蛋白(TP)、向蛋白(Alb)、总胆红素(T-Bil)、谷丙转氨酶(GPT)、谷草转氨酶(GOT)、乳酸脱氢酶(LDH)6个项目的测量值相关性较好,相关系数(r)均大于0.975,配对t检验分析结果差异均无统计学意义。批内精密度均在CLIA'88允许误差1/4内,批间精密度均在CLIA'88允许误差1/3内,符合临床实验室质控要求。结论SPOTCHEM-4410分析仪在航行舰船环境中应用时检测结果准确、稳定,有较好的抗干扰能力。Objective To evaluate the clinical application analysis performance of the SPOTCHEM-4410 analysator in medical transport ship environment. Methods Detection results in hospital laboratory and medical ship during navigation were measured,then accuracy and precision of the SPOTCHEM-4410 analysator were evaluated, respectively. Results There were positive correlation of total protein (TP), albumin (Alb), bilirubin total (T-Bil), glutamate-pyruvate transaminase (GPT), glutamate-oxalacetic transaminase ( GOT ) and lactate dehydrogenase(LDH) in the tests before and after navigation (r〉0.975, P〉0.05 ). The error of measurement in medical ship during navigation were lower than 1/4 orl/3 of the CLIA'88 permissible errors, which was permitted by precision requirement. Conclusions The SPOTCHEM-4410 analysator is accurate, stable and has strong resistance against interference.

关 键 词:干试剂化学分析仪 卫生船舶 临床分析性能 

分 类 号:R686[医药卫生—骨科学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象