个体化充气止血带充气压力的研究  

Investigation of the Best Charged Pressure of Pneumatic Tourniquet in Individualized Using

在线阅读下载全文

作  者:卜玲[1] 田斌斌[1] 陈顺芳[1] 顾卫华[1] 李志红[1] 

机构地区:[1]泰州市中医院,江苏泰州225300

出  处:《泰州职业技术学院学报》2008年第3期69-70,78,共3页Journal of Taizhou Polytechnic College

摘  要:目的探讨充气止血带个体化使用时的适宜充气压力。方法选择200例四肢骨折病人,随机分为两组,分别采用个体化的充气压力和常规充气压力,比较两种充气压力下的止血效果及止血带副损伤发生情况。结果两组止血效果均良好,而个体化充气压力止血带副损伤发生率远低于常规压力下。结论以记录气压值+50mmHg(上肢)、记录气压值+100mmHg(下肢)设定为电动止血带的充气压力。Objective To investigate the best charged pressure of pneumatic tourniquet in individualized using. Methods 200 limb fracture patients were divided into two groups. Individualized charged pressure and routine pressure were used in two groups, The hemostatic result and side reactions were compared within the two groups. Results The hemostatic results of two groups were favorable. In the group of individualized pressure patients of the individualized had more side reactions than that of the routine group. Conclusion Taking the recorded pressure plus 50mmHg(upper limbs) and recorded pressure plus 100mmHg(lower limbs) is suitable pressure of pneumatic tourniquet.

关 键 词:充气止血带 个体化 止血效果 副损伤 

分 类 号:R473[医药卫生—护理学]

 